IJSR International Journal of Scientific Research 2277 - 8179 Indian Society for Health and Advanced Research ijsr-9-4-24851 Original Research Paper Comparative accuracy of apex locators in detecting lateral canals, extent of perforating internal root resorption and open apex- An ex-vivo study Afzal Dr. Vineeta Nikhil Dr. April 2020 9 4 01 02 ABSTRACT

Aim: This study compared the accuracy of Root ZX mini and CanalPro in determination of working length in open apex, location of lateral canals at different levels and determination of extent of perforating internal root resorption. Methods: Hundred decoronated single rooted teeth with single canal were divided into three groups– simulated lateral canals at 3, 5 and 8 mm from the apex, simulated perforating internal root resorption and simulated open apex. Actual length was measured by visual method and electronic length of samples embedded in alginate model. Results: The accuracy of both the apex locators in EL determination was significantly higher in apical lateral canals compared to middle and coronal canals and in apical extent of internal resorption as compared to coronal extent (P<0.05). The accuracy of both apex locators was more at ± 1 mm when compared to ± 0.5 mm. Conclusion: Both EAL’s are accurate in working length determination in simulated clinical scenarios.