Volume : IV, Issue : VI, June - 2015
COMPARISON OF RADIATION DOSE IN BRAIN IMAGING BETWEEN 64 SLICE AND 16 SLICE CT
Jawad Islam T, Sushil Yadav, Rahul P. Kotian
Abstract :
Aim:Comparison of radiation dose in ain imaging between 16 and 64 slice CT. Materials and Methods:Sampling was done by convenience sampling method considering inclusion and exclusion criteria and a sample of 80 was included. 40 were scanned in 64 slice CT and 40 were scanned in 16 slice CT. Effective dose was computed from DLP using the equation; E=k x DLPk=0.0021 (According to European guidelines). Statistical analysis was done using Mann–Whitney U test.Result:The 16 slice CT machine is delivering more radiation dose compared to 64 slice CT for ain imaging with statistically significant difference.Conclusion:In the context of ain CT imaging using the same software from the same manufacturer, 64CT examinations provide gives a lower ED when compared with 16CT examinations
Keywords :
Article:
Download PDF
DOI : 10.36106/ijsr
Cite This Article:
Jawad Islam T, Sushil Yadav, Rahul P. Kotian Comparison of Radiation Dose in Brain
Imaging Between 64 Slice And 16 Slice Ct
International Journal of Scientific Research, Vol : 4, Issue : 6 June 2015
Number of Downloads : 1152
References :
Jawad Islam T, Sushil Yadav, Rahul P. Kotian Comparison of Radiation Dose in Brain Imaging Between 64 Slice And 16 Slice Ct International Journal of Scientific Research, Vol : 4, Issue : 6 June 2015