Volume : III, Issue : II, February - 2014

Automated Erythrocyte Sedimentation Rate Analyser V/S The Westergren’s Manual Method In Measurement Of Erythrocyte Sedimentation Rate: A Comparative Study

Dr. Gauravi Dhruva, Dr. Amit Agravat, Dr. Mahesh Kakadiya, Dr. Hemant Pansuriya

Abstract :

MicroSED-10® (Electra Lab,Italy) is automated method for measurement of erythrocyte sedimentation rate (ESR). The aim of our study was to compare the ESR values by MicroSED-10® against the standard Westergren’s method. Study was conducted in outpatient department of pathology at P.D.U. MEDICAL COLLEGE & HOSPITAL-RAJKOT on 209 patients. The samples taken were as per the recommendations charted out by International Council for Standardization in Haematology (ICSH). Bland and Altman statistical analysis was applied for evaluation. The analysis revealed a low degree of agreement between the manual and automated method especially for higher ESR values[>25mm/hour],mean difference~2.58±9.17(95% limits of agreement,-15.39 to 20.55) as compare to lower ESR value[<25mm/hour],mean difference~1.22±1.90(95% limits of agreement,-2.50 to 4.94) & Over all whole ESR data analysis shows, mean difference~2.37±8.34(95% limits agreement,?13.94 to 18.68) for 1 hour. The fully automated system MicroSED-10® for ESR measurement tends to underestimate the manual ESR readings especially at higher ESR values.

Keywords :

Article: Download PDF   DOI : 10.36106/ijsr  

Cite This Article:

Dr. Gauravi Dhruva, Dr. Amit Agravat, Dr. Mahesh Kakadiya, Dr. Hemant Pansuriya Automated Erythrocyte Sedimentation Rate Analyser V/S The Westergren’s Manual Method In Measurement Of Erythrocyte Sedimentation Rate: A Comparative Study International Journal of Scientific Research, Vol.III, Issue.II February 2014


Number of Downloads : 2320


References :